Model Number 16-02-95
Device Problems Loss of Power (1475); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/11/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device.He identified a poor contact of power supply conversion cable plug.He replaced the conversion cable and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t shutted down during use.The procedure could be completed by reconnecting the power outlet and turning the power off and on and the issue re-occurred multiple times.The engineer of the hospital said that the shape of the power plug looks a little strange and bent.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the serial number of the device has been added to the dedicated section d.4 and the h.4 has been filled in accordingly.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9443053
MDR Text Key196429435
Report Number9611109-2019-00957
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-95
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/11/2019
Initial Date FDA Received12/10/2019
Supplement Dates Manufacturer Received01/08/2020
Supplement Dates FDA Received02/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-