• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ290
Device Problem Mechanical Problem (1384)
Patient Problem Injury (2348)
Event Date 09/07/2019
Event Type  Injury  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the (b)(6) that the duodenal mucosal injury and erosion occurred because the up angle was not up to standard during the diagnostic procedure using the subject device.The patient was suggested to take oral medication for treatment.Other information such as outcome of the patient was not provided.
 
Manufacturer Narrative
This supplemental report is being submitted to additional information.Olympus medical systems corp.(omsc) reviewed the manufacture history (dhr) of the device and confirmed no irregularity.The exact cause could not be determined, but there was the possibility that the user applied the excessive bending and tensile stress to the subject device, it might cause the stretch of the angle wire, consequently the up angle might not be up to standard.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9443496
MDR Text Key185501429
Report Number8010047-2019-04214
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberGIF-HQ290
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/10/2019
Supplement Dates Manufacturer Received03/11/2020
Supplement Dates FDA Received03/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-