• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK STATUS+ Back to Search Results
Catalog Number 10379675
Device Problem False Negative Result (1225)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2019
Event Type  malfunction  
Manufacturer Narrative
Proper technique and maintenance were reviewed with the customer and determined to have been followed.The customer no longer had reagent or patient sample available for further investigation.A review of the complaint database has shown no related complaints for this lot.Customer has had no further discrepancies.The instrument is operational, running hcg tests.The cause of this event is unknown.
 
Event Description
The customer reported a false negative urine hcg results on the clinitek status+ when compared to three home pregnancy tests and a serum quantitative hcg.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key9447714
MDR Text Key174174720
Report Number3002637618-2019-00129
Device Sequence Number1
Product Code JHI
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 12/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10379675
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/10/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-