DEXCOM, INC. DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number N/A
Device Problem Unintended Application Program Shut Down (4032)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that an unexpected cgm app shut-off occurred.It was indicated that the operating system for the smart device was not a supported system, which is misuse of the device.Data was evaluated and the allegation was confirmed.The probable cause was determined to be the mobile device updated its operating system which closed the app.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 GLUCOSE PROGRAM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9452650
MDR Text Key173138021
Report Number3004753838-2019-105034
Device Sequence Number1
Product Code QDK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberSW11785
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/13/2019
Initial Date FDA Received12/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
-
-