ZIMMER MANUFACTURING B.V. BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Insufficient Information (3190)
Patient Problems No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
Event Date 11/27/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the bone screw fractured while fixing the g7 shell.The tip side of the screw remained in the screw hole of the shell.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D4: udi#: (b)(4).The event was confirmed with product received.-visual evaluation identified that the threaded portion of the screw had fractured, and the fractured piece was not returned.Surface damage was observed on the spherical radius and shank.No other damages were identified.Sem analysis of the screw's fracture surface showed suspected crack initiation and exit areas, and indications of a torsional overload fracture.Sheared ductile overload dimples were identified near the center of the fracture surface.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 30 MM LENGTH
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key9452939
MDR Text Key190835071
Report Number0002648920-2019-00889
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K934765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00625006530
Device Lot Number64146346
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/11/2019
Supplement Dates Manufacturer Received04/28/2020
Supplement Dates FDA Received04/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN G7 SHELL
Patient Outcome(s) Other;
-
-