Model Number N/A |
Device Problems
Fracture (1260); Insufficient Information (3190)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
Event Date 11/27/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Foreign report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the bone screw fractured while fixing the g7 shell.The tip side of the screw remained in the screw hole of the shell.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.D4: udi#: (b)(4).The event was confirmed with product received.-visual evaluation identified that the threaded portion of the screw had fractured, and the fractured piece was not returned.Surface damage was observed on the spherical radius and shank.No other damages were identified.Sem analysis of the screw's fracture surface showed suspected crack initiation and exit areas, and indications of a torsional overload fracture.Sheared ductile overload dimples were identified near the center of the fracture surface.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|