Investigation summary: one sample was received.It has opened packaging flow wrap.It has the plunger rod-rubber stopper, the tip cap, saline solution.The luer tip is bent.This occurs when the syringe is misplaced in the sterilizer tray.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: this is the 1st complaint for lot # 9227890 for this type of defect or symptom.There was no documentation for this type of defect during the entire production run of this batch.Root cause description: this occurs when the syringe is misplaced in the sterilizer tray.
|
It was reported that a bent syringe was found before use with a 10 ml bd posiflush¿ sp pre-filled flush syringe nacl 0.9%.The following information was provided by the initial reporter, "when the outer plastic was removed from the 10ml normal saline syringe flush, it was noticed by the rn that the syringe was disfigured.The top of the syringe was bent out of shape.".
|