• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND COMPANY 10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9%

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND COMPANY 10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9% Back to Search Results
Catalog Number 306575
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: one sample was received.It has opened packaging flow wrap.It has the plunger rod-rubber stopper, the tip cap, saline solution.The luer tip is bent.This occurs when the syringe is misplaced in the sterilizer tray.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: this is the 1st complaint for lot # 9227890 for this type of defect or symptom.There was no documentation for this type of defect during the entire production run of this batch.Root cause description: this occurs when the syringe is misplaced in the sterilizer tray.
 
Event Description
It was reported that a bent syringe was found before use with a 10 ml bd posiflush¿ sp pre-filled flush syringe nacl 0.9%.The following information was provided by the initial reporter, "when the outer plastic was removed from the 10ml normal saline syringe flush, it was noticed by the rn that the syringe was disfigured.The top of the syringe was bent out of shape.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 ML BD POSIFLUSH¿ SP PRE-FILLED FLUSH SYRINGE NACL 0.9%
Type of Device
FLUSH
Manufacturer (Section D)
BECTON DICKINSON AND COMPANY
2153 12th avenue
columbus NE 68601
Manufacturer (Section G)
BECTON DICKINSON AND COMPANY
2153 12th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key9456085
MDR Text Key194204871
Report Number1911916-2019-01304
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2022
Device Catalogue Number306575
Device Lot Number9227890
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2019
Initial Date FDA Received12/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-