• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY Back to Search Results
Model Number 022765
Device Problems Malposition of Device (2616); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device 13 of 17.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event Description
The end user reported that 17 wafers from two boxes with the same lot number had off-centered starter hole.Out of these, 12 wafers were unused.The remaining 5 wafers were used and the end user experienced reduced wear time with these.No photo is available at this time.
 
Manufacturer Narrative
Batch record review: the batch record review supports that there were no discrepancies related to the issue reported.This batch was manufactured following the applicable procedures, all machine and testing parameters were in accordance with the applicable procedures and specifications, the testing results were found satisfactory and no nonconformance related to the malfunction code in analysis was noted.The line clearance was executed per dr-sop-0094 ¿procedimiento de despeje de línea y chequeo de seguridad (line clearance procedure and safety check)¿, the results were documented, and no issues were identified during the manufacturing process.The materials comply with the correct parameters, procedure instruction (pi), ds, mt, tm and the results are satisfactory.Photograph, video and/or physical sample evaluation: no photographs associated with this case were received and no unused return sample was expected.Executive summary of the non-conformance (nc): based on the results of the preliminary investigation, the batch record reviews were performed, and no discrepancies were found.A revision of the improvements made to the process was performed, and actions have been implemented for this type of incidents.In addition, an upgrade of the guard robot arm, disc collar and pouch assembly process was performed.These improvements performed to the guard machine, contribute to the reduction of possible defects related to off center.The last batch received was manufactured prior to the implementation of these improvements.Furthermore, a complaints search conducted shows that no adverse trend was identified from january 2019 to august 2020 for this failure mode.This failure mode will keep monitoring for track and trending.For this reason, no additional investigation is needed.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Type of Device
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key9456865
MDR Text Key177601120
Report Number9618003-2019-17379
Device Sequence Number1
Product Code EZQ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number022765
Device Lot Number9G04454
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/12/2019
Supplement Dates Manufacturer Received08/28/2023
Supplement Dates FDA Received09/25/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient SexFemale
-
-