• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER Back to Search Results
Model Number 119216M
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the temp sensing catheter did not produce accurate readings.The complainant stated the patient was normothermic.However, despite checking/adjusting the equipment, the reading was 94.5f.No medical intervention was reported.
 
Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause is "sensor wire too weak or thermistor chip too weak".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "for urological use only single use only do not resterilize.Note: compatible with appropriate 400-series temperature monitors.Interchangeability + 0.2oc at 37oc.Caution: as with all temperature probes: in the presence of rf energy sources: local heating, temperature errors, and probe damage may occur.In medical use, unplug the temperature-sensing catheter at the extension cable before activating electrosurgical or other types of direct coupled rf energy sources.Do not stretch catheter.This will cause repositioning of probe.Do not use stylet.This will cause stretching of catheter.Valve type: use luer slip syringe.Do not use needle.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile: unless package is opened or damaged.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Recommended inflation capacities 8 fr.And 10 fr.(3cc balloon): use 3.5cc sterile water 12 fr.(5cc balloon): use 5.5cc sterile water 14 fr.And larger (5cc balloon): use 10cc sterile water do not exceed recommended capacities.Caution: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Event Description
It was reported that the temp sensing catheter did not produce accurate readings.The complainant stated the patient was normothermic.However, despite checking/adjusting the equipment, the reading was 94.5f.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER
Type of Device
TEMP SENSING FOLEY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9465613
MDR Text Key178748119
Report Number1018233-2019-07946
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741046308
UDI-Public(01)00801741046308
Combination Product (y/n)N
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model Number119216M
Device Catalogue Number119216M
Device Lot NumberNGDU1907
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/21/2019
Initial Date FDA Received12/13/2019
Supplement Dates Manufacturer Received12/18/2019
Supplement Dates FDA Received12/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-