Section h10: (h3) per capa: 2017-12, the risk documentation was lacking specific failures for this type of defect for these instruments.In addition, the surgical technique was not providing information to the user of the instrument that reaming off axis can damage or break the instruments.Additionally, the surgical technique did not provide adequate instruction to the user.In this case, due to the tip on the device being found fractured on the back table prior to use in the reported surgery, it is likely that the tip was damaged during a previous surgery and subsequent reprocessing and handling resulted in the propagation and ultimate fracture of the pilot tip.(h9) z-2663-2017, z-2664-2017, z-2665-2017, z-2666-2017, z-2667-2017, z-2668-2017.
|