The customer did not return any product.The suspect medical device has been updated.Applicable fields of sections d and g were updated.The (b)(6) analyzer serial number was (b)(6).Section h9 was updated.During internal testing, roche determined the limits of detection (lod) for protein, nitrite, leukocytes, and erythrocytes on the (b)(6) urine analyzer with chemstrip® 5 ob, chemstrip 7, chemstrip 10 md, and chemstrip 10 ua test strips were higher than what is listed in their respective test strip method sheets.This can lead to false negative results on the (b)(6) urine analyzer for the four affected parameters.Roche has provided customers with the following instructions and work arounds: chemstrip 5 ob and chemstrip 7 test strips should only be read visually.They can no longer be read on the (b)(6)urine analyzer.Chemstrip 10 md or chemstrip 10 ua test strips can be used with the (b)(6) urine analyzer; however, a negative result for any one of the four affected parameters (i.E., protein, nitrite, leukocytes, and erythrocytes) on the (b)(6) urine analyzer must be repeated with a new test strip that is read visually.
|