H.10: an increasing trend has been identified for loose shaft encoder issue on centrifugal pump 5 (cp5).This triggered a detailed investigation which revealed that in case of loose shaft encoder on cp5 system the control of the pump speed is not impacted.Indeed, the shaft angle encoder movements are limited by mechanical constraint of the housing/circuit boards.A loose shaft encoder will only result in less than 0,5 seconds of unresponsiveness which is the time needed by the encoder to reach the housing.Thus, the pump speed can still be controlled by the user through the knob.Based on the above the reported event has been re-assessed as non reportable.
|