• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.He adjusted the knob and the issue was solved.Functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 double head pump speed control knob was loose after the procedure.There was no report of patient injury.
 
Manufacturer Narrative
H.10: an increasing trend has been identified for loose shaft encoder issue on centrifugal pump 5 (cp5).This triggered a detailed investigation which revealed that in case of loose shaft encoder on cp5 system the control of the pump speed is not impacted.Indeed, the shaft angle encoder movements are limited by mechanical constraint of the housing/circuit boards.A loose shaft encoder will only result in less than 0,5 seconds of unresponsiveness which is the time needed by the encoder to reach the housing.Thus, the pump speed can still be controlled by the user through the knob.Based on the above the reported event has been re-assessed as non reportable.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9467468
MDR Text Key194320879
Report Number9611109-2019-00968
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2019
Initial Date FDA Received12/13/2019
Supplement Dates Manufacturer Received04/16/2020
Supplement Dates FDA Received05/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-