STRYKER ORTHOPAEDICS-MAHWAH LEFT HEMIPELVIS REPL SIZE H TRIDENT ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number C-M110-1-400
Device Problems Off-Label Use (1494); Insufficient Information (3190)
Patient Problems Scar Tissue (2060); Injury (2348); No Known Impact Or Consequence To Patient (2692)
Event Date 11/17/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There has been 1 other similar event for the lot referenced.This was for the same device in the same patient for an intraoperative event.
 
Event Description
This pi is to address the second surgery (b)(6) 2019.Primary procedure, left custom heavy pelvis.It was reported that after implanting the replacement pelvis and two axsos screws, the surgeon noticed excessive bleeding and had to remove the devices to find the source.The surgeon decided to abort the procedure and closed the patient.Post-operatively, the patient is stable.The surgeon reported his belief that in removing scar tissue (patient diagnosed with ewings sarcoma and had a previous procedure to resect bone), he may have hit a vessel which caused the bleeding.Rep confirmed there are no allegations against the implants.Update (b)(6) 2019: hospital re-sterilized the implant and cut off parts to modify for the patient.The surgery was then completed successfully with no further information provided by the surgeon and hospital.
 
Manufacturer Narrative
An event regarding a revision involving left hemipelvis repl size h trident id was reported.Conclusion: an initial attempt to implant this device was made on (b)(6) 2019.It was reported that after implanting the replacement pelvis and two axsos screws, the surgeon noticed excessive bleeding and had to remove the devices to find the source.The surgeon decided to abort the procedure and closed the patient (pi (b)(4)).Post-operatively, the patient was stable.The surgeon reported his belief that in removing scar tissue (patient diagnosed with ewings sarcoma and had a previous procedure to resect bone), he may have hit a vessel which caused the bleeding.On (b)(6) 2019, the surgeon modified the implant, re-sterilized it and successfully implanted it on (b)(6) 2019.No further information was provided.Based on this provided information it has been determined that this event is associated with an off-label application.A review of the device manufacturing records indicated qin (b)(4), was packaged with the device at the time of manufacture.The router does not indicate the version of the package insert.A review of the most recent version of the document, rev aa, noted the following: sterilization these products have been sterilized by either gamma radiation or hydrogen peroxide.Refer to the package label for the sterilization method.If the package is opened, but the product is not used, the prosthesis must not be resterilized and must therefore be discarded or returned to the supplier.Single use devices cannot be explanted and subsequently reimplanted as the physical forces exerted by these actions may compromise the physical integrity, dimensions and/or surface finishes of the devices.Also, sterility cannot be assured for reused devices as cleaning and re-sterilization procedures have not been verified.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
This pi is to address the second surgery on (b)(6) 2019.Primary procedure, left custom heavy pelvis.It was reported that after implanting the replacement pelvis and two axsos screws, the surgeon noticed excessive bleeding and had to remove the devices to find the source.The surgeon decided to abort the procedure and closed the patient.Post-operatively, the patient is stable.The surgeon reported his belief that in removing scar tissue (patient diagnosed with ewings sarcoma and had a previous procedure to resect bone), he may have hit a vessel which caused the bleeding.Rep confirmed there are no allegations against the implants.Update 19/nov/2019: hospital re-sterilized the implant and cut off parts to modify for the patient.The surgery was then completed successfully with no further information provided by the surgeon and hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEFT HEMIPELVIS REPL SIZE H TRIDENT ID
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9469465
MDR Text Key176556459
Report Number0002249697-2019-04025
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberC-M110-1-400
Device Lot Number9K59PD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/17/2019
Initial Date FDA Received12/13/2019
Supplement Dates Manufacturer Received02/28/2020
Supplement Dates FDA Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age15 YR
Patient Weight143
-
-