KARL STORZ SE & CO. KG INNER SHEATH WITH CERAMIC BEAK Back to Search Results
Model Number 27050CA
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/29/2019
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by the manufacturer, karl storz in (b)(6).As per the evaluation: the broken ceramic tip is cracked and shows slight discoloration due to fire damage is visible at the distal end.The slight discoloration indicates too high/too long exposure to heat.The ceramic tip has been mechanically overloaded - most likely dropped and is what pre-damaged the ceramic tip.An indicator for this is the crack formation at the ceramic tip.Damages like this happen if the ceramic insert is hit against hard objects or edges.During intended use, there is no possibility to apply forces to achieve the defects.The warning notices with the corresponding pictures in the ifu indicate this.
 
Event Description
As per a vigilance report filed with the (b)(6) competent authority by our parent company in (b)(6) , during a planned procedure, resection of the prostate, our instruments were used.Right at the beginning of the procedure, the ceramic tip of the inner shaft became detached.After a long, intensive search, the broken tip was recovered.Due to surgical extension, the patient was unnecessarily exposed to anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH WITH CERAMIC BEAK
Type of Device
INNER SHEATH WITH CERAMIC BEAK
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM  
Manufacturer Contact
susie chen
2151 e. grand avenue
el segundo, CA 90245-5017
4242188201
MDR Report Key9469715
MDR Text Key188081284
Report Number9610617-2019-00118
Device Sequence Number1
Product Code FBO
UDI-Device Identifier04048551077412
UDI-Public4048551077412
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K943668
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27050CA
Device Catalogue Number27050CA
Device Lot NumberRS05
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/29/2019
Initial Date FDA Received12/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-