ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Information (3190)
Event Date 11/15/2019
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "defib pad short," "check pads," and "poor pad contact" messages.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation evaluated the device and the device performed to specification.The device was recertified and returned to the customer.Review of copies of the device activity log printouts were provided and evaluated.The log shows that in one occurrence the shock button was pressed and/or held down by the user before the device completed the charging cycle and another occurence that the charge button was pressed more times than was necessary (the customer indicated that the device did emit the charging sounds).There is no evidence in the log to indicate that the device was not functioning as intended and the logs showed several instances of successfully delivered shocks.The clinical data or device activity log files were not available as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key9469718
MDR Text Key170786370
Report Number1220908-2019-03887
Device Sequence Number1
Product Code DRT
UDI-Device Identifier00847946017521
UDI-Public00847946017521
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/22/2019
Initial Date FDA Received12/13/2019
Supplement Dates Manufacturer Received11/22/2019
Supplement Dates FDA Received01/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-