Model Number SN6AT5
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/21/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
A facility representative reported that during a cataract surgery with intraocular lens (iol) implantation, a defect was found on the center of the optic.The iol was removed from the eye and replaced with a different iol to complete the procedure.Additional information has been requested.
 
Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information was provided in h.3., h.6.And h.10.Evaluation summary: the product was returned.Solution is dried on the lens.Haptic and optic damage were observed.All product and batch history records are quality reviewed prior to product release.A qualified associated product was indicated.An unspecified handpiece was indicated.It is unknown if qualified product combinations were used.Two viscoelastics were indicated, only one is qualified for the lens/cartridge combination indicated.The reported optic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key9470263
MDR Text Key176768226
Report Number1119421-2019-02152
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 02/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Model NumberSN6AT5
Device Catalogue NumberSN6AT5.180
Device Lot Number12370791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2019
Initial Date Manufacturer Received 11/22/2019
Initial Date FDA Received12/13/2019
Supplement Dates Manufacturer Received12/16/2019
01/22/2020
Supplement Dates FDA Received01/09/2020
02/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MONARCH III D CARTRIDGES.; UNSPECIFIED DUOVISC.; UNSPECIFIED MONARCH IOL DELIVERY SYST.
Patient Age88 YR
-
-