• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR IMPLANT 54; ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR IMPLANT 54; ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 999804754
Device Problems Nonstandard Device (1420); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Test Result (2695); No Code Available (3191)
Event Date 11/04/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).No 510(k) number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr revision.Asr xl: left hip.Reason for revision: component loosening.Rcf notes elevated chromium/cobalt, esr and crp levels, mri shows periprosthetic lucency suggestive of loosening.Doi: (b)(6) 2006 - dor: (b)(6) 2019.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes heen able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not bsynthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: h6 product complaint # (b)(4).Investigation summary: the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa- 001226.Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR ACETABULAR IMPLANT 54
Type of Device
ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key9476044
MDR Text Key185849287
Report Number1818910-2019-123239
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number999804754
Device Lot Number1176862
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/23/2019
Initial Date FDA Received12/16/2019
Supplement Dates Manufacturer Received01/23/2020
01/30/2020
Supplement Dates FDA Received01/27/2020
02/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-