APOLLO ENDOSURGERY, INC. BIB SYSTEM INTRAGASTRIC BALLOON Back to Search Results
Model Number B-40800
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Code Available (3191)
Event Date 11/12/2019
Event Type  malfunction  
Manufacturer Narrative
Apollo has not received the product at this time.Therefore no analysis or testing has been done.A review of the device labeling notes the following: the current bib¿ system intragastric balloon directions for use (dfu) addresses the known and anticipated potential event of "broken device" as follows: warnings and precautions: proper positioning of the placement catheter assembly and the bib¿ system balloon within the stomach is necessary to allow proper inflation.Lodging of the balloon in the esophageal opening during inflation may cause injury and/or device rupture.The bib¿ system balloon is composed of soft silicone elastomer and is easily damaged by instruments or sharp objects.The balloon must be handled only with gloved hands and with the instruments recommended in this document.Note: if the balloon becomes separated from the sheath prior to placement, do not attempt to use the balloon or reinsert the balloon into the sheath.Note: during the filling process the fill tube must remain slack.If the fill tube is under tension during the intubation process, the fill tube may dislodge from the balloon, preventing further balloon deployment.Warning: rapid fill rates will generate high pressure which can damage the bib¿ system valve or cause premature detachment.
 
Event Description
Health professional reported via product field notes (pfn), "the patient didn't notice any satiety and the balloon was split.Another balloon was placed.".
 
Manufacturer Narrative
Supplement #1-medwatch sent to the fda on 17/mar/2020.The reporter noted that the device will not be returned.Additional information: d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIB SYSTEM INTRAGASTRIC BALLOON
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key9476687
MDR Text Key183690079
Report Number3006722112-2019-00198
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2020
Device Model NumberB-40800
Device Catalogue NumberB-40800
Device Lot NumberAF02091
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/21/2019
Initial Date FDA Received12/16/2019
Supplement Dates Manufacturer Received11/21/2019
Supplement Dates FDA Received03/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age39 YR
-
-