APOLLO ENDOSURGERY, INC. ORBERA 365 INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-50012
Device Problem Leak/Splash (1354)
Patient Problems Discharge (2225); Patient Problem/Medical Problem (2688)
Event Date 11/24/2019
Event Type  malfunction  
Manufacturer Narrative
Apollo has not received the product at this time.Therefore no analysis or testing has been done.A review of the device labeling notes the following: the current orbera365¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "bowel complications", "deflation", "early removal" as follows: precautions: antiemetics, antispasmodic, and anticholinergic drugs may be prescribed to lessen the early placement symptoms such as nausea, vomiting, and abdominal pain.Patients will need to immediately contact their physician for any severe or unusual symptoms.Placement of the balloon within the stomach produces an expected and predictable reaction characterized most commonly by a feeling of heaviness in the abdomen, nausea and vomiting, gastroesophageal reflux, belching, esophagitis, heartburn, diarrhea and, at times, abdominal, back or epigastric pain and cramping.Food digestion may be slowed during this adjustment period.These symptoms can be treated with antiemetic, antispasmodic, and anticholinergic medications.Typically the stomach acclimates to the presence of the device within the first 2 weeks.In order to prevent or ameliorate the symptoms most frequently experienced during the adjustment period, it is recommended that the physician use proton pump inhibitors (ppis), antiemetics, antispasmodics, and anticholinergic medications prophylactically (before orbera¿ placement).Patients should be advised to immediately contact their physician for any unusually severe or worsening symptoms.The physiological response of the patient to the presence of orbera365¿ may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response.Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients need to be evaluated and the device removed at or within 12 months of placement.Clinical data does not exist to support use of an individual orbera365¿ beyond 12 months.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications that may result from the use of this product include the risks associated with the medications and methods utilized in the endoscopic procedure, the risks associated with any endoscopic procedure, the risks associated with the orbera intragastric balloon specifically, and the risks associated with the patient's degree of intolerance to a foreign object placed in the stomach.The physiological response of the patient to the presence of the orbera365 system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response.Deflated devices should be removed promptly.Complications: possible complications of the use of the orbera365 system include: balloon deflation and subsequent removal.
 
Event Description
Reported complaint: "the patient defecated a deflated intragastric balloon.Upon further inspection, a silicone layer rupture about 2-3cm long was discovered.The event took place at the patients home, no medical staff was present.The patient has sent photos of the ruptured device via email and the device itself via post service.The patient refuses a reimplantation.This is the 8th orbera balloon migration event in our clinical experience - further information may be acquired from (b)(6).".
 
Manufacturer Narrative
Device evaluation summary: the device was returned to the apollo device analysis laboratory on (b)(6) 2019.A balloon was returned discolored with a large hole present.There were no other components returned with the device.Due to the large hole present in the shell, functional evaluation could not be conducted.Observation of the hole shows smooth edges with the edges rolling inward with significant discoloration present on the balloon shell.
 
Manufacturer Narrative
Supplement #1: the device was returned to the apollo device analysis laboratory on 13/dec/2019.Analysis of the device is ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA 365 INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
MDR Report Key9476694
MDR Text Key183682880
Report Number3006722112-2019-00203
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/17/2020
Device Model NumberB-50012
Device Catalogue NumberB-50012
Device Lot NumberAF02209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/16/2019
Supplement Dates Manufacturer Received12/04/2019
12/04/2019
Supplement Dates FDA Received12/20/2019
12/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient Weight86
-
-