• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE ORTHOANCHOR; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE ORTHOANCHOR; PLATE Back to Search Results
Model Number 01-301-06-09
Device Problem Break (1069)
Patient Problem Impaired Healing (2378)
Event Date 11/19/2019
Event Type  Injury  
Manufacturer Narrative
An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Manufacturer Narrative
(b)(4).Reference exemption number e2017029.
 
Event Description
It was reported a plate broke and was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE ORTHOANCHOR
Type of Device
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
Manufacturer (Section G)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM   78570
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key9476815
MDR Text Key184390147
Report Number9610905-2019-00258
Device Sequence Number1
Product Code DZE
UDI-Device Identifier00888118005816
UDI-Public(01)00888118005816
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040891
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number01-301-06-09
Device Lot Number33265220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/18/2020
Event Location Hospital
Date Report to Manufacturer11/19/2019
Initial Date Manufacturer Received 11/19/2019
Initial Date FDA Received12/16/2019
Supplement Dates Manufacturer Received02/18/2020
Supplement Dates FDA Received03/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age17 YR
-
-