• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated an issue with diagnostic collection counters.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable cardiac monitor (icm) exhibited invalid counters.The device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9480183
MDR Text Key173520476
Report Number9614453-2019-04267
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2017
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-