• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 3-4 11MM; PRSTHSISKNEEPTLLOFMROTIBLSMICNSTRINDCMNTDPLYMRMTLPOLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 3-4 11MM; PRSTHSISKNEEPTLLOFMROTIBLSMICNSTRINDCMNTDPLYMRMTLPOLYMR Back to Search Results
Catalog Number 71453212
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Injury (2348); Joint Swelling (2356)
Event Date 12/18/2018
Event Type  Injury  
Event Description
Primary surgery performed on 2016.It was reported that on (b)(6) 2018 patient felt an unusual sensation when standing up from a seated position, the joint felt loose and had a clicking noise.Patient knee and lower leg started to swell and pain increased.A scan was performed at (b)(6) hospital, where they diagnosed a burst bakers cyst.In (b)(6), patient contacted her surgeon; a joint x-ray was performed, doctor diagnosed a broken and displaced insert.A revision surgery was performed to correct the issue.During the operation it was discovered that the post on the insert had sheared off and the metal parts had worn on each other leaving significant metallic debris around the joint and metallosis.Since then, the second knee has taken much longer to recover and patient still needs the aid of crutches to walk significant distances and requires the nightly use of diclofenac 3% gel.
 
Manufacturer Narrative
Additional information received by the manufacturer has identified that this event should be re-evaluated for mdr reporting.The new information states that this event was already reported on mdr, case no: 1020279-2019-00802, therefore, it was determined that this case does not meet the threshold for reporting.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 3-4 11MM
Type of Device
PRSTHSISKNEEPTLLOFMROTIBLSMICNSTRINDCMNTDPLYMRMTLPOLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9481030
MDR Text Key178150806
Report Number1020279-2019-04427
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556034170
UDI-Public00885556034170
Combination Product (y/n)N
PMA/PMN Number
K071071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71453212
Device Lot Number11KM07793
Initial Date Manufacturer Received 11/21/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received11/21/2019
Supplement Dates FDA Received12/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-