U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH; TROCAR Back to Search Results
Model Number G407209
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Great Vessel Perforation (2152)
Event Date 12/12/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
Related manufacturing ref: 3005334138-2019-00907.During an atrial fibrillation ablation procedure, a perforation of the aorta occurred during transseptal puncture.The puncture was too anterior and a perforation was confirmed vis fluoroscopy.The patient was transferred to cardiac surgery for repair and is in stable condition.There were not any performance issue with any abbott devices.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported perforation remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRK-1¿ XS TRANSSEPTAL NEEDLE, 71 CM LENGTH
Type of Device
TROCAR
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key9483579
MDR Text Key173167240
Report Number3008452825-2019-00641
Device Sequence Number1
Product Code DRC
UDI-Device Identifier05414734205184
UDI-Public05414734205184
Combination Product (y/n)N
PMA/PMN Number
K122587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model NumberG407209
Device Catalogue NumberG407209
Device Lot Number7121697
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received02/03/2020
Supplement Dates FDA Received02/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SWARTZ¿ BRAIDED INTRODUCER, SL0
Patient Outcome(s) Required Intervention;
-
-