• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2019
Event Type  malfunction  
Manufacturer Narrative
Product analysis:analysis could not confirm the customer comment.Output connectors were not broken.Could not fully insert the pacing cable.Atrial output connector was contaminated inside and out.Hanger was cracked.Keypad window was scratched.It was contaminated under rate and menu knobs.Both output nuts were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) had broken atrial and ventricular ports.The caller was advised to return the epg to service; the epg remains in use at this time.No patient complications have been reported as a result of this event.The epg was returned for service.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) had broken atrial and ventricular ports.The caller was advised to return the programmer to service; the programmer remains in use at this time.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9484368
MDR Text Key175553432
Report Number3004593495-2019-01315
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00763000162443
UDI-Public00763000162443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 01/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/17/2019
Supplement Dates Manufacturer Received12/23/2019
01/10/2020
01/14/2020
Supplement Dates FDA Received01/08/2020
01/13/2020
01/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-