MAUDE Adverse Event Report: THERAKOS INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number CLXUSA
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/18/2019
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h339 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h339 shows no trends.Trends were reviewed for complaint category, centrifuge bowl leak/break.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer contacted mallinckrodt to report that they experienced a centrifuge bowl leak/break with their cellex photopheresis kit ("kit") during an extracorporeal photophereis (ecp) treatment.The customer stated that approximately 149 mls of whole blood was processed at the time the leak occurred.Customer stated that the treatment was aborted and the patient was stable.The customer did not return any product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key9488139
MDR Text Key209012422
Report Number2523595-2019-00148
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)H339(17)210601
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2021
Device Model NumberCLXUSA
Device Catalogue NumberCLXUSA
Device Lot NumberH339
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2019
Initial Date FDA Received12/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient Weight56
-
-