Model Number 8888135191
Device Problem Product Quality Problem (1506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, after the tube was fixed, the ring of the fixed wing came out of the tube without damage (the ring fixed on the tube body separated from the body).It was stated that the suture wing was not broken, the stiches held, and there was nothing unusual observed on the device.The catheter was not repaired.There was no leak.Lodophor was the cleaning agent used.Tego was not utilized and there was no luer adapter issue.The catheter was replaced to resolve the issue.There was no reported patient injury.
 
Manufacturer Narrative
Additional information: g4, h3, h6.H3 evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.The visual inspection of the returned photo noted: the suture wing is attached to the patient.The catheter is not in the photograph.Without the physical device functional is precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key9488461
MDR Text Key194118005
Report Number3009211636-2019-00274
Device Sequence Number1
Product Code MPB
UDI-Device Identifier10884521006447
UDI-Public10884521006447
Combination Product (y/n)N
PMA/PMN Number
K030209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/29/2023
Device Model Number8888135191
Device Catalogue Number8888135191
Device Lot Number1831100168
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/18/2019
Supplement Dates Manufacturer Received02/12/2020
Supplement Dates FDA Received02/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-