The reporter indicated the surgeon found a black dot-like foreign matter on a 13.2mm vicm5_13.2 implantable collamer lens, -06.00 diopter, when the lens was removed from the vial.The lens was not used or implanted.The backup lens was implanted and the problem was resolved.The cause of the event was due to the device.
|
H3: device evaluation: the lens was returned dry in a micro-centrifuge vial.Visual inspection found no visible damage to the lens.H6: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim#: (b)(4).
|