Model Number 0570-0338 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/14/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The glidescope gvm video monitor was returned to verathon for evaluation.A verathon technical service representative evaluated the returned monitor where they were unable to duplicate the reported image issue.It was noted that when the unit arrived, it would not power on due to the battery being depleted.Once the battery was charged, the unit would power on and function as expected.After observing proper device operation through functional and performance testing, the device was certified and returned to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope gvm video monitor, the image would disappear intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
|