VERATHON MEDICAL ULC GLIDESCOPE GVM MONITOR; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0338
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2019
Event Type  malfunction  
Manufacturer Narrative
The glidescope gvm video monitor was returned to verathon for evaluation.A verathon technical service representative evaluated the returned monitor where they were unable to duplicate the reported image issue.It was noted that when the unit arrived, it would not power on due to the battery being depleted.Once the battery was charged, the unit would power on and function as expected.After observing proper device operation through functional and performance testing, the device was certified and returned to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, using a glidescope gvm video monitor, the image would disappear intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE GVM MONITOR
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key9492348
MDR Text Key188577812
Report Number9615393-2019-00132
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0338
Device Catalogue Number0570-0338
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2019
Initial Date Manufacturer Received 11/20/2019
Initial Date FDA Received12/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-