The customer reported that during a patient procedure, using a glidescope spectrum smart cable, the image would disappear intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
D10, g4, g7, h2, h3, h6, h10.The glidescope core smart cable was returned to verathon for evaluation.A verathon technical service representative evaluated the returned cable where they were unable to verify the reported issue.When connected to a test blade and monitor, the device would display an image that was stable and clear with good color rendition.Since the customer was provided with a replacement device, the returned cable was scrapped.No further investigation is required at this time.Verathon will continue to monitor for trends.
|