• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD F&P SLEEPSTYLE AUTO CPAP; BZD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD F&P SLEEPSTYLE AUTO CPAP; BZD Back to Search Results
Model Number SPSAAN
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint f&p sleepstyle series cpap was received at fisher & paykel healthcare in (b)(4) and was visually inspected internally and externally.Results: visual inspection showed that one pin had detached from the mains inlet socket.Conclusion: the reported incident was traced to an issue in the assembly process of the supplied mains inlet connector component.The supplier of the component was notified and they have made changes to the assembly process.As part of our ongoing product improvement initiatives, we implemented a gauge test which identifies and rejects any potentially faulty mains inlet sockets prior to assembly into the sleepstyle.The subject sleepstyle was manufactured prior to implementation of these measures.Our user instructions that accompany the f&p sleepstyle state the following: do not use if the device, power cord or accessories are damaged, deformed, or cracked.Do not pull on the power cord as it may become damaged.Turn the device off at the power supply, then remove the power cord from the rear of the device.
 
Event Description
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative an issue with s sleepstyle unit.Upon device evaluation at fisher & paykel healthcare (f&p) in (b)(4) on 25 november 2019, it was found that a pin from the mains inlet socket had come apart.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F&P SLEEPSTYLE AUTO CPAP
Type of Device
BZD
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
8007923912
MDR Report Key9492801
MDR Text Key181943692
Report Number9611451-2019-01234
Device Sequence Number1
Product Code BZD
UDI-Device Identifier09420012437068
UDI-Public(01)09420012437068(10)2100588068(11)180919
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K173193
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSPSAAN
Device Catalogue NumberSPSAAN
Device Lot Number2100592542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2019
Initial Date Manufacturer Received 11/25/2019
Initial Date FDA Received12/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-