Model Number 801188 |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) was unable to duplicate the reported complaint.He replaced the oxygen (o2) sensor.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the unit prompted the user to recalibrate the electronic patient gas system (epgs) following initial calibration.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) observed the oxygen (o2) sensor to pass calibration 10 consecutive times.The unit operated to the manufacturer's specifications.Per data log analysis, it was confirmed that the initial calibration failed displaying an 'o2 sensor out of range' at 7:01:45 am.The user tried to recalibrate and received the same error at 8:42:05 am and at 8:43:45am.At 13:03:19 calibration was started and successfully passed at 13:05:07.Between the failed gas calibrations the fraction of oxygen (fio2) and flow was changed several times and there were o2 sensor and blender disagree errors.
|
|
Event Description
|
Additional information received that the reported issue occurred during set-up.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|