TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801188
Device Problem Calibration Problem (2890)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) was unable to duplicate the reported complaint.He replaced the oxygen (o2) sensor.The unit operated to the manufacturer's specifications.The suspect device was returned to the manufacturer for further evaluation.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the unit prompted the user to recalibrate the electronic patient gas system (epgs) following initial calibration.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed the oxygen (o2) sensor to pass calibration 10 consecutive times.The unit operated to the manufacturer's specifications.Per data log analysis, it was confirmed that the initial calibration failed displaying an 'o2 sensor out of range' at 7:01:45 am.The user tried to recalibrate and received the same error at 8:42:05 am and at 8:43:45am.At 13:03:19 calibration was started and successfully passed at 13:05:07.Between the failed gas calibrations the fraction of oxygen (fio2) and flow was changed several times and there were o2 sensor and blender disagree errors.
 
Event Description
Additional information received that the reported issue occurred during set-up.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9494504
MDR Text Key203712333
Report Number1828100-2019-00675
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)080416
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/19/2019
Supplement Dates Manufacturer Received01/09/2020
01/29/2020
Supplement Dates FDA Received01/29/2020
02/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-