Model Number 3100 A
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2019
Event Type  Injury  
Manufacturer Narrative
This unit was field serviced by a vyaire medical authorized service technician.The service technician was able to verify the customer's reported issue during testing and evaluated.The service technician calibrated the unit and noticed the circuit on the unit was wet.It likely that the water got inside of the transducer.The service technician will replace the transducer to address the reported issue.
 
Event Description
It was reported to vyaire medical that the map was fluctuating while in use on a patient.There was no patient harm associated with this reported event.The patient was placed on another device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
stanley tan
22745 savi ranch pkwy
yorba linda, CA 92887
7149193324
MDR Report Key9498199
MDR Text Key173274306
Report Number2021710-2019-11180
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20050627
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2019
Initial Date FDA Received12/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-