MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 12/19/2019.
 
Event Description
It was reported that the diagnostic code related to ventilator restart due to anomalies occurred during periodic maintenance.There was no patient involvement.The manufacturer¿s international service technician evaluated the ventilator and confirmed the reported problem.The service technician replaced the central processing unit printed circuit board assembly and the problem was resolved.
 
Manufacturer Narrative
G4: 17mar2020.B4: 19mar2020.The replaced central processing unit (cpu) printed circuit board assembly (pcba) was returned for failure analysis.The cpu pcba was tested and no failures were identified.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9500125
MDR Text Key181195895
Report Number2031642-2019-10928
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2019
Initial Date FDA Received12/19/2019
Supplement Dates Manufacturer Received12/13/2019
Supplement Dates FDA Received03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-