Model Number 60-02-60
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).Video and pictures confirming the reported event have been provided.The cp5 control panel and the drive unit were requested by livanova for a dedicated investigation.A serial read-out of the cp5 control panel was performed: the reported error was confirmed.It was found that the centrifuge plug and the processor board were not right connected.However, despite the hardware deviation detected, during the functional test no deviations could be detected and the centrifuge worked without issues.
 
Event Description
Livanova received a report that two different faults in motor controller, related to the centrifugal pump unit, were displayed on the centrifugal pump 5 (cp5) control panel when the cable connecting the drive unit to the panel was slightly moved.Moreover, it could be possible to detect a change in the rpms when the cable was moved.The issue occurred during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9502003
MDR Text Key196573407
Report Number9611109-2019-00981
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/11/2019
Initial Date FDA Received12/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-