Model Number 50-80-00 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).The investigation is ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 mast roller pump displayed a motor control error during priming.The pump was turned off and on and it worked properly again.The device was used also in the following days with no issues.There was no report of patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: a review of the dhr could not identify any deviations or nonconformities relevant to the issue.A field service technician was dispatched to the facility to investigate the device and identified that the cooling fan of the pump was completely blocked by dust.The fan was cleaned and the issue solved.The most likely root cause of the reported event is the dirty fan which could have led to an overheating of the electronic components and to bad connection issues.
|
|