• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 50-80-00
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/22/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).The investigation is ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump displayed a motor control error during priming.The pump was turned off and on and it worked properly again.The device was used also in the following days with no issues.There was no report of patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: a review of the dhr could not identify any deviations or nonconformities relevant to the issue.A field service technician was dispatched to the facility to investigate the device and identified that the cooling fan of the pump was completely blocked by dust.The fan was cleaned and the issue solved.The most likely root cause of the reported event is the dirty fan which could have led to an overheating of the electronic components and to bad connection issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
MDR Report Key9502005
MDR Text Key197334883
Report Number9611109-2019-00979
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number50-80-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/25/2019
Initial Date FDA Received12/20/2019
Supplement Dates Manufacturer Received12/17/2019
Supplement Dates FDA Received01/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-