MEDTRONIC NAVIGATION, INC NAVIGATION STEALTH STATION S8 EM ENT SYSTEM; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9735669 |
Device Problem
Imprecision (1307)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
No devices were returned to the manufacturer for analysis.Other relevant device(s) are: product id #: 9735762, software version #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system used during a functional endoscopic sinus surgery (fess) procedure.It was reported that after registering the patient with a 1.8 metric, they were touching known landmarks and the surgeon was satisfied with the accuracy.They then used the malleable suction and microdebrider and the surgeon stated he felt that he was 4-5 mm inaccurate posterior.They then added two more points and lowered the metric to 1.6.They then touched known landmarks and the surgeon felt accurate until the malleable suction was brought in and again the surgeon stated that he felt inaccurate 4-5 mm posterior.There was a delay of less than 1 hour.No patient impact was correlated with this event.
|
|
Manufacturer Narrative
|
H3:a software analysis was initiated.However, the software evaluation found that a probable cause was unable to be determined since the behavior cannot be replicated.Codes associated with the software: fdr 3221, fdc 4315.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
D2 correction: product code change from haw to pgw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|