Device evaluation: device was returned with visible traces of blood and examined in the biohazard area of the lab.As received, the connector was observed to be slightly detached from the cannula body.A leak test was performed on the cannula and leakage was observed from the connector hub section.No other visual damage, contamination, or other abnormalities were found to the device.Additional manufacturer narrative: the device was returned to edwards for evaluation.Customer complaint of cracked femii1018a was confirmed with assessment.The device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.As reported, the cracked cannula was found during use.Inspections are in place at edwards during receiving and packaging to mitigate device damage.Damage due to shipping or clinical use cannot be ruled out.Neither a supplier nor an edwards¿ defect can be confirmed.A definitive root cause could not be identified.
|