Model Number 5392
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 11/20/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) top rate knob would not correctly change the rate.It was noted that the knob worked part time, but mostly it does not respond when the knob is turned.The epg was returned for repair.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that the rate knob would not adjust correctly and the encoder flex is out of specification.It was also determined at analysis that the ventricular output connector was broken, hanger was cracked, ventricular pace pulse noise and atrial pace pulse noise, the main printed circuit board (pcb) was out of specification, the battery drawer sticks and the main label was damaged due to customers device tracker.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis: failure analysis: confirmed customer complaint - encoder aging - fracture of internal contact terminals caused encoder malfunction.No defect found for atrial and ventricular pulse output noise.The high measurement of a-v pulse output noise on the automated test console is due to noise from the automated test console combining with the device noise.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key9510730
MDR Text Key175652292
Report Number3004593495-2019-01345
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Initial Date Manufacturer Received 12/16/2019
Initial Date FDA Received12/23/2019
Supplement Dates Manufacturer Received01/07/2020
07/30/2020
Supplement Dates FDA Received01/24/2020
08/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-