• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Intermittent Continuity (1121)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 23dec2019.
 
Event Description
The customer reported during pm (preventive maintenance) service they found the touchscreen with intermittent operation.The field service engineer (fse) confirmed intermittent touchscreen operation and found the unit has intermittent touchscreen operation during parameter changes.The fse performed touchscreen calibration, but the issue was not resolved.The device was not in clinical use at the time the issue was discovered, therefore, there was no patient or user harm reported.The fse replaced the touchscreen to resolve the issue.The fse powered on the unit and the touchscreen was operational.
 
Manufacturer Narrative
G4: 24mar2020 b4: (b)(6)2020.Touchscreen assembly was received for evaluation.There were no signs of damage or contamination during visual inspection.After testing, it was determined that the ul_lr and ur_ll resistance and resistance ratio were out of specification, which caused the failure.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9511788
MDR Text Key182143411
Report Number2031642-2019-10938
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/23/2019
Supplement Dates Manufacturer Received12/06/2019
Supplement Dates FDA Received03/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-