Model Number D-AVHD-DF16
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2019
Event Type  malfunction  
Manufacturer Narrative
One advisor¿ hd grid mapping catheter, sensor enabled¿ was received for investigation.Investigation revealed electrode 17 was displaced and the conductor wire had been fractured proximal to the solder joint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the displaced electrode and open circuit is consistent with damage during use.
 
Event Description
This event is being reported based on the analysis of the returned device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL (AFD-PLYMOUTH)
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key9512778
MDR Text Key190957787
Report Number3005334138-2019-00928
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberD-AVHD-DF16
Device Lot Number6967353
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/16/2019
Initial Date FDA Received12/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-