Model Number AUT75
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/30/2017
Event Type  malfunction  
Manufacturer Narrative
This peeled/bubbled paint was believed to be caused by improper maintenance such as periodic disinfection and excessive use of cleaning agents.
 
Event Description
This mdr is being reported at this time as part of an internal review of past complaints.Due to the incident being in the past, the information that can be obtained from the initial reporter is limited.On december 15, 2017, dai-ichi shomei received information that on the rotating part, paint was peeled off and had some bubbles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AURORA 3 SERIES
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
DKK DAI-ICHI SHOMEI CO., LTD.
32-26, sakashita 1-chome
itabashi-ku, 174-0 043
JA  174-0043
Manufacturer (Section G)
DKK DAI-ICHI SHOMEI CO., LTD.
32-26, sakashita 1-chome
itabashi-ku, 174-0 043
JA   174-0043
Manufacturer Contact
kenneth block
800 e. campbell road
suite 202
richardson, TX 75081
9724809554
MDR Report Key9513571
MDR Text Key190691742
Report Number3006437518-2019-00010
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberAUT75
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/15/2017
Initial Date FDA Received12/23/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-