MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
|
Back to Search Results |
|
Model Number 9732719 |
Device Problem
Imprecision (1307)
|
Patient Problem
Radiation Exposure, Unintended (3164)
|
Event Date 12/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device unique identifier (udi) is unavailable.No devices were returned to the manufacturer for analysis.Device manufacturing date is unavailable.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging device being used intra/peri-operatively of a sacroiliac and thoracolumbar procedure.It was reported that the surgeon found out that, when using the imaging system, the navigation system was off by 4-5 mm "proud." the surgeon was not happy and used another navigation system to completed the procedure.When using another imaging system.It was found when checking for accuracy, it was 2-3 mm deeper then where the surgeon was at.The site completed the procedure while continuing to check accuracy.It was successful after confirming screw placement and with the final imaging scan before inserting rods.There was an 18-minute delay to the procedure and no impact on patient outcome.
|
|
Manufacturer Narrative
|
Product id: bi-500-01093 software version: 3.1.7.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: a software analysis was initiated.The analysis found that the reported issue was unrelated to the software of the imaging system.The software functioned as designed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|