• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION¿ ENT NAVIGATION SYSTEM; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION¿ ENT NAVIGATION SYSTEM; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733560XOM
Device Problem Human-Device Interface Problem (2949)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/24/2019
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 97 33467, software version: 2.3.No devices were returned to the manufacturer for analysis.Device manufacturing date is unavailable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a fusion system used during a functional endoscopic sinus surgery (fess) procedure.It was reported that during a clinical case the site was having issues registering the patient.The site had rebooted a few times and had gone into the register portion of the software and stated that they were only able to get the 3.4mm quadcut to show up for the registration.The site was unable to verify any other instruments on the system.After speaking again with the site it was known that the tracker that was being used was not an ent tracker.Navigation was aborted.Troubleshooting was performed which includes reboot the system and check the instrumentation being used for the procedure.Patient was present.Unknown delay in the case.The case was resolved on the call.On 2019-dec-26 additional information received: patient information received, length of surgical time was less than 1 hour.
 
Manufacturer Narrative
H3: a software analysis was initiated.However, the software evaluation determined that per case description, the behavior described is the intended behavior of the software.Software is functioning as designed.Codes associated with the software: fdr 213, fdc 67 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION¿ ENT NAVIGATION SYSTEM
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9523036
MDR Text Key184851801
Report Number1723170-2019-06180
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733560XOM
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/24/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received02/20/2020
Supplement Dates FDA Received02/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age42 YR
Patient Weight112
-
-