• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM Back to Search Results
Catalog Number UNK HIP FEMORAL STEM
Device Problems Degraded (1153); Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Necrosis (1971); Pain (1994); Loss of Range of Motion (2032); Distress (2329); Test Result (2695); No Code Available (3191); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/06/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised to address elevated metal ion level.Doi: unknown; dor: (b)(6) 2019; unknown affected side.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 health effect - clinical code: appropriate term / code not available (e2402) used to capture the blood heavy metal increased and bone injury.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.E3 initial reporter occupation: lawyer.
 
Event Description
After review of medical records patient was revised to addressed painful right total hip arthroplasty.After splitting the gluteus maximus there was significant amount of metallosis along the posterior aspect of the hip which had infiltrated the region of the trochanteric bursa.A large bursal sac posterior and lateral to the hip which was excised and there was cloudy fluid resting within.Posterior pseudocapsule infiltrated with metallosis and necrotic tissue.The large metal head was removed and noted mild degree of corrosion at the trunnion.Acetabular component was removed with minimal bone loss and also within the pubis.Femoral component is well fixed in approximately 25 to 35 degrees of anteversion.Doi: (b)(6) 2009.Dor: (b)(6) 2019 right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Litigation records received.Records allege osteolysis, loss of mobility, adverse reaction to metal wear debris, emotional distress, and economic loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL STEM
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9525687
MDR Text Key175076314
Report Number1818910-2019-124762
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL STEM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/22/2020
06/29/2021
09/15/2021
Supplement Dates FDA Received01/24/2020
07/14/2021
09/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ASR UNI FEMORAL IMPL SIZE 43; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL SLEEVE ASR
Patient Outcome(s) Required Intervention;
Patient Age53 YR
-
-