• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Mechanical Problem (1384)
Patient Problem No Information (3190)
Event Date 12/17/2019
Event Type  malfunction  
Manufacturer Narrative
Vyaire complaint number: (b)(4).Any additional information provided by the customer will be included in a follow up report.At this time, vyaire has not received the suspect device/component for evaluation.
 
Event Description
It was reported to vyaire that the 3100 a experienced fluctuations with mean airway pressure (map) and amplitude settings.At this time, it is unknown if there was any patient involvement associated with the reported issue.Tests/lab data, including dates.
 
Manufacturer Narrative
Vyaire complaint #: (b)(4).Device evaluation: g6, h2, h3, h6, h10 results of investigation: a vyaire medical service representative (fsr) evaluated the device onsite and verified the reported problem.Fsr found fluid in the bias flow line.The fluid was removed, and the device ran normally.Device passed performance testing.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
MDR Report Key9526700
MDR Text Key189477806
Report Number2021710-2019-11205
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20150422
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/17/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received12/17/2019
Supplement Dates FDA Received02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-