• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ARTHRO RASP 20 UPBEND *EA; BONE RASP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ARTHRO RASP 20 UPBEND *EA; BONE RASP Back to Search Results
Model Number 214619
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).The lot number is unknown.
 
Event Description
It was reported by affiliate via email that during an unknown procedure the surgeon noticed that rasp shaft of a rasp 20° up bend, was bent.The surgery was completed by using the same bent instrument.No surgical delay or patient consequence reported.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary : according to the information provided, it was reported that during surgeon noticed that rasp shaft was bent.The complaint device is not being returned, multiples attempts for product return were made with no response, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.  given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.  since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.  at this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This rentrt has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that the device was bent from leveraging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
BONE RASP
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9526886
MDR Text Key208140770
Report Number1221934-2019-60185
Device Sequence Number1
Product Code NBH
UDI-Device Identifier10886705004027
UDI-Public10886705004027
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214619
Device Catalogue Number214619
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/13/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/13/2020
01/28/2020
Supplement Dates FDA Received01/15/2020
01/29/2020
Patient Sequence Number1
-
-