• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION COMPACT; EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9 735638, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system that was used outside of procedure.It was reported that when the site attempted to turn on the system, they got the green power light and the manufacturer splash screen, but then the screen went fully blank.When they tried to reboot, they go the same screen again before the system powered off.There was no patient present at the time of the event.
 
Event Description
Additional information received from a manufacturer representative indicated no further actions were taken.The system was no longer under a service agreement and would be returned.
 
Manufacturer Narrative
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: 9735638, version #: 1.0 ; product id: 9735602, serial/lot #: (b)(6), udi#: (b)(4); product id: 9731203, serial/lot #: (b)(6), udi#: (b)(4).H3: hardware analysis of the system computer confirmed the reported behavior; graphics card malfunction.The computer booted to a splash screen, then it went blank.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Hardware analysis of the returned field generator determined the enclosure to be damage.However, the field generator functioned as intended; tracked normally through the duration of testing.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Software analysis determined the system software was functioning as intended.Conclusion code 4307 and result code 4203 pertain to system computer and field generator analysis.Conclusion code 67 and result code 213 pertain to system computer analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION COMPACT
Type of Device
EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9527269
MDR Text Key184851274
Report Number1723170-2019-06198
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169672956
UDI-Public00643169672956
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735602
Device Catalogue Number9735602
Device Lot Number1839496
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2020
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/30/2020
Supplement Dates FDA Received02/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-