• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190766
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem Cramp(s) (2193)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility biomedical technician (biomed) reported that a fresenius hemodialysis (hd) machine removed too much weight during a patient's hd treatment.It was unknown what the machine was set to remove and how much was actually removed.There were no reported diagnostic messages or audible alarms.It was reported the patient was cramping.During recirculation, the machine was set to remove 100 ml and it removed 150 ml.Upon follow-up, the biomed confirmed that there was no patient injury, adverse events, or medical intervention required as a result of the reported event.The biomed stated that the machine was returned to service after the ultrafiltration (uf) pump was replaced.The biomed could not confirm if the old uf pump was available to be returned to the manufacturer for evaluation.The treatment details were unknown.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
MDR Report Key9527447
MDR Text Key178333780
Report Number2937457-2019-03657
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (y/n)N
PMA/PMN Number
K121341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 01/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number190766
Device Catalogue Number190766
Was Device Available for Evaluation? No
Device Age MO
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/17/2020
Supplement Dates FDA Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FRESENIUS BLOODLINES.; FRESENIUS DIALYZER.
-
-