MEDTRONIC NAVIGATION, INC (LITTLETON) BASE OARM BI70000028120 SYS 120V; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
|
Back to Search Results |
|
Model Number BI70000028120 |
Device Problem
Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient involved.A medtronic representative went to the site to test the equipment.The system failed the mechanical inspection due to application error on mobile viewing station stating that the image acquisition system config file could not load.Technical service instructed the msb rep, to reload config file from usb backup over the phone.This resolved the issue and the system perform as intended.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system used outside of procedure.It was reported that prior to a case when booting the system the mobile viewing station (mvs) displayed an error stating the config file was corrupt.The msb rep reported the issue to the fse rep.The rep had reload the config file, and he will complete the system checkout.No patient present at the time of event.Additional information received on dec 11 2019: application error on mvs stating that the ias config file could not load.Technical services instructed the msb rep, to reload config file from usb backup over the phone.This resolved issue.Performed site visit and could not duplicate symptom.Reloaded 3.2.1 software.System checkout performed.
|
|