• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) BASE OARM BI70000028120 SYS 120V; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) BASE OARM BI70000028120 SYS 120V; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI70000028120
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Patient Involvement (2645)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
No patient involved.A medtronic representative went to the site to test the equipment.The system failed the mechanical inspection due to application error on mobile viewing station stating that the image acquisition system config file could not load.Technical service instructed the msb rep, to reload config file from usb backup over the phone.This resolved the issue and the system perform as intended.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system used outside of procedure.It was reported that prior to a case when booting the system the mobile viewing station (mvs) displayed an error stating the config file was corrupt.The msb rep reported the issue to the fse rep.The rep had reload the config file, and he will complete the system checkout.No patient present at the time of event.Additional information received on dec 11 2019: application error on mvs stating that the ias config file could not load.Technical services instructed the msb rep, to reload config file from usb backup over the phone.This resolved issue.Performed site visit and could not duplicate symptom.Reloaded 3.2.1 software.System checkout performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE OARM BI70000028120 SYS 120V
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9527652
MDR Text Key180035651
Report Number3004785967-2019-02198
Device Sequence Number1
Product Code OXO
UDI-Device Identifier00643169353428
UDI-Public00643169353428
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000028120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/09/2019
Initial Date FDA Received12/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-