One actual sample and one companion sample were received for evaluation.The other actual sample was not received and therefore, could not be evaluated.Visual inspection of both samples did not identify any abnormalities that could have contributed to the reported condition.All components are correctly placed and according to specifications.Functional tests which included pressure and clear passage under water tests were performed with no issues noted.The actual sample and the companion sample both met product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|