• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8537
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that two (2) units of clearlink system continu-flo solution sets leaked; further described as once connected to the bag of intravenous (iv) fluid it leaked where the drip chamber connects to the tubing.This issue was identified after priming before connecting to the patient.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
One actual sample and one companion sample were received for evaluation.The other actual sample was not received and therefore, could not be evaluated.Visual inspection of both samples did not identify any abnormalities that could have contributed to the reported condition.All components are correctly placed and according to specifications.Functional tests which included pressure and clear passage under water tests were performed with no issues noted.The actual sample and the companion sample both met product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9527819
MDR Text Key173041484
Report Number1416980-2019-07216
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048970
UDI-Public(01)00085412048970
Combination Product (y/n)N
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8537
Device Lot NumberR19I28092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2020
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/27/2019
Supplement Dates Manufacturer Received01/13/2020
Supplement Dates FDA Received01/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-