• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP® SYSTEM CLIP DELIVERY SYSTEM; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP® SYSTEM CLIP DELIVERY SYSTEM; VALVE REPAIR Back to Search Results
Catalog Number CDS0602-XTR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Tissue Damage (2104)
Event Date 12/11/2019
Event Type  Injury  
Manufacturer Narrative
Exemption number e2019001.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of mitral valve injury (tissue damage) as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.Based on the information available, the reported patient effect of tissue damage appears to be due to procedural conditions.There is no indication of a product quality issue with respect to manufacture, design or labeling.The other two mitraclip devices are filed under a separate medwatch report number.
 
Event Description
This is filed to report tissue damage.It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4+.The first clip delivery system (cds 90926u451) was advanced to the mitral valve.The final arm angle (efaa) test failed, the clip opened to 50-60 degrees.The sixth efaa attempt was successful; however, the decision was made to not implant the clip and the cds was removed.A second cds (90926u452) was advanced and the clip was placed.However, during the efaa test the clip slightly opened 20-25 degrees.Leaflet insertion was still very good, therefore the clip was implanted.After deployment, the clip opened more than prior to deployment, but remained secure on the leaflets.A third clip (90926u496) was advanced to further reduce mr; after the leaflets were grasped, the posterior leaflet tore.The clip was implanted; mr was reduced to 3-4.The patient was hemodynamically stable and was extubated without any problems.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP® SYSTEM CLIP DELIVERY SYSTEM
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9527929
MDR Text Key174901986
Report Number2024168-2019-15050
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/27/2020
Device Catalogue NumberCDS0602-XTR
Device Lot Number90926U496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received12/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-